Chip makes pH measurements easier
Researchers have developed a new technology that makes pH measurements more robust, simple and reliable.
Instead of the traditional, often error-prone reference electrodes, a durable chip is now used. This chip can be stored dry, is pressure-resistant and easily integrated into compact devices.
Initial test kits are already available for rapid use in medical, biological, agricultural and environmental applications.
Precise pH measurement is a challenge in chemical sensor technology. To date, conventional silver/silver chloride/potassium chloride electrodes have been used.
Although these electrodes provide a stable reference potential, they are highly prone to errors. Changes in electrolyte concentration or a clogged or dried-out membrane can make the readings unreliable.
While the pH-sensitive glass electrode has already been successfully replaced by robust chips called ion-sensitive field-effect transistors (ISFETs), replacing the sensitive reference electrode has recently been achieved at Fraunhofer IPMS with a reference ISFET (REFET).
“Our ISFETs are reliable, stable and durable. They overcome common problems of conventional reference electrodes, such as clogged or charged diaphragms, drift due to concentration changes, and contamination of the measurement solution,” explained Dr Olaf Hild, division director of the Chemical Sensors and Systems department at Fraunhofer IPMS. This makes them particularly suitable for mobile or integrated measurement systems.
In this technology, developed by Fraunhofer IPMS, a second ISFET is used with a standard pH ISFET instead of a sensitive reference electrode. This reference ISFET has a significantly lower pH slope of 20 mV/pH at 25°C, compared to the usual 59 mV/pH according to the Nernst equation.
The ISFETs are coated with thin layers of niobium pentoxide (Nb2O5) or tantalum pentoxide (Ta2O5). These materials provide the sensors with high stability, ease of handling and good storage properties. The two ISFETs operate together via an auxiliary electrode to reliably calculate the pH value from the measured electrical signals.
“Unfortunately, we did not develop the ISFET-REFET concept. It was created in the 1980s by Professor Piet Bergveld, the inventor of the ISFET, and his colleagues,” Hild said. “However, the concept has so far not been adopted commercially.”
The new components aim to change this. Initially pH measurement will be possible in the range of approximately pH 4 to pH 8, which meets the typical measurement requirements for many applications in biology, medicine, agriculture and environmental monitoring.
Currently, two 5×5 mm chips are required for proper wetting, but their size can be reduced upon customer's request. In the future, the both chips are planned to be combined into a single chip of the same size, with integrated temperature measurement.
“Long-term measurements after a 2- or 3-point calibration require precise control of sensor drift. The drift can be compensated for with the help of the control electronics," said electronics developer Hans-Georg Dallmann, describing how the control system works.
This establishes the roadmap for further development at Fraunhofer IPMS: expanding the pH range of the REFET through improved sensor layers, reducing REFET sensor drift, and integrating both functions on a single chip with temperature measurement.
“Although much work still lies ahead, we are pleased to already offer test kits and to present them to the professional audience at Analytica 2026,” Hild said.
Individual appointments at the event can be arranged in advance via the Fraunhofer IPMS website.
UNSW tracks groundwater recharging: study
Researchers are looking at when Australia's groundwater recharges and how climate change is...
Aussie-made remote telemetry unit tech expands into NZ
CGI has signed a new reseller agreement with Landis+Gyr New Zealand to expand the availability of...
$10m project to safeguard WA water supplies
Construction is set to begin on a 7.4 km pipeline linking Sovereign Hill to Guilderton in the...